XEI Scientific, Inc.
RF Plasma Cleaning Systems for Electron Microscopes
and High Vacuum Systems

The EVACTRON® Anti-Contaminator and De-Contaminator
Stops Artifacts and Removes Hydrocarbons and Organics.

Updated August 2007

XPS EVALUATION OF SAMPLES SURFACE CLEANED BY
THE XEI EVACTRON®
Scott D. Walck***, Brian R. Strohmeier***,
Edward G. Goralski***, and Ronald A. Vane

***PPG Industries, Inc., Pittsburgh, PA 15238

Abstract of paper presented at M&M 2001
Long Beach, CA August 2001

With the advent of field emission gun equipped SEMs and TEMs with clean vacuum systems, it has been shown repeatedly that the surface cleanliness of samples exposed to the small, high current density electron beam available in these instruments is critical with respect to controlling hydrocarbon contamination under the electron beam. In previous studies of dedicated plasma cleaners for TEM samples by one of the authors (SDW), it was shown that surface analytical techniques are the best methods for determining the effectiveness of such cleaners in removing hydrocarbons from the samples' surfaces.1, 2 In the current study, some of the experiments that were used in these previous studies were repeated using the XEI Evactron® 3 system in order to compare the relative effectiveness of this system for surface cleaning of EM samples.

Figure 1 shows a picture of the sample introduction chamber of a VG Scientific ESCALAB Mk II system (XPS) with the Evactron® in place. The Evactron® system consists of an RF head unit that attaches to the vacuum chamber with an adaptable flange, an RF impedance matching unit, and an RF generator/vacuum controller. Although a 100% O2 system can be used and is recommended, for safety reasons, a 20% O2, 80% Ar by volume gas mixture was selected because of the hydrocarbon oil used in the mechanical pump. Cleaning was also performed using laboratory air. The major difference between the dedicated plasma cleaners and the Evactron® configuration is that the sample is not immersed in the plasma but is exposed to the activated oxygen species in the downstream flow.

Samples of Si, Ti, and Ni were first solvent cleaned (SC) and then intentionally contaminated by either a low temperature melting wax (LT-Wax) in acetone solution or exposure to the exhaust of a mechanical pump with a hydrocarbon oil (Oil), and then plasma cleaned (PC) with the Evactron® system with increments of 20 min of time. XPS analysis was done on the SC samples, intentionally contaminated samples, and Evactron® cleaned samples. One important difference in the present study compared to the earlier studies was that the sample was processed in the sample introduction chamber of the VG XPS system and thus was not exposed to ambient air prior to analysis.

Table 1 gives the XPS concentration results of the samples intentionally contaminated cleaned with the Evactron® with air and Ar O2 for 20 min. The Ar O2 mixture typically reduced the carbon concentration better than when air was used. Although the Evactron® has significantly decreased the hydrocarbon level on the surfaces, it has not completely eliminated all of it. In comparison, the dedicated plasma cleaners were able to remove all of the contaminates with a 20 min treatment.1 Increasing the time to 40 min and 60 min further decreases the oil, but again did not eliminate it (not shown). Figure 2 shows the C 1s peak for the oil contaminated Si sample after a 20 min PC treatment. The spectrum from the oil contaminated sample shows a typical aliphatic carbon peak characteristic of C-C and C-H bonds. After the PC treatment, an additional carbonyl peak and carboxyl shoulder are present indicating the oxidation of the hydrocarbon species. Figure 3 shows the Si 2p spectra of a Si SC sample before and after a 20 min PC treatment. It can be seen that there are little relative changes between the oxide and the metal peaks indicating that the Si surface is not oxidizing. In contrast, in the previous study involving dedicated plasma cleaners, oxidation of the solvent cleaned Si surface was quite apparent and could also be detected in the TEM.1

In conclusion, the Evactron® unit does decrease the hydrocarbons present on sample surfaces. With properly prepared and handled samples, it would most likely remove modest amounts of contaminates even with the reduced O2 partial pressure used in this study. It was unfortunate that 100% O2 could not be used because it probably would have produced better results. It is also evident from the variable geometry possible with an Evactron® that care should be taken with the design parameters if it used as a sample cleaner, such as distance and position of the sample relative to the RF head and pump ports. The lack of oxidation of the Si surface suggests that it is a more gentle cleaning solution for contamination problems than the dedicated plasma cleaners.

References

1J T Grant, S D Walck, et al., MRS Proceedings, Vol. 480, pp. 49-71, 1997.

2 S P Roberts, N J Zaluzec, S D Walck, J T Grant, MRS Proceedings, Vol. 480, pp. 127-136, 1997.

3 R A Vane, US Patent Number 6,105,589, Aug 22, 2000.