XEI Scientific, Inc.
RF Plasma Cleaning Systems for Electron Microscopes
and High Vacuum Systems

The EVACTRON® Anti-Contaminator and De-Contaminator
Stops Artifacts and Removes Hydrocarbons and Organics.

Updated August 2007

Reducing Oil Backstreaming in
Electron Microscopes

by Ronald Vane, XEI Scientific

INTRODUCTION

Of the sources of contamination, vacuum pump oil backstreaming is the easiest to attack. The original version following report was written before I founded XEI Scientific to provide Electron Microscopists more satisfactory methods stopping contamination. Because of this the SEM-CLEAN Nitrogen purge system and Evactron Decontaminator system are not mentioned as solutions. Dry Pump technolgy has also advanced since 1990, and there are many more models than mentioned below that are available from various manufacturers. -R.V.

Some types of Electron Microscopes are standard equipped with Oil Diffusion Pumps for High Vacuum and Oil Sealed Rotary pumps for backing and roughing. Under some conditions oil backstreaming can be a problem to the user. This oil can cause sample contamination, oil films on detector windows, dirty apertures, and other problems for the microscopist. A number of methods can be used to attack this problem with a choice based on the cost, maintenance schedule, reliability, and achievable vacuum.

BACKSTREAMING

To understand the solutions to the backstreaming problem it is important to understand the types of flow found in vacuum systems. In rough vacuums above about 10-2 mbar ( 1 Pa) we find viscous flow conditions. In viscous flow the gas acts like a fluid and moves from areas of high pressure to those of low pressure. Each molecule moves with the molecules around it. If the forward flow rate is faster than the backwards diffusion rate there can be no backstreaming in viscous flow conditions.

At high vacuums below about 10-5 mbar ( 10-3 Pa) we find molecular flow conditions. The mean free path between molecular collisions becomes long enough that each molecule acts independently of the movement of surrounding molecules. For this reason high vacuum pumps act by capturing the molecules that come in and prevent them from returning to the vacuum chamber. To pump efficiently under molecular flow conditions wide pump openings are needed to capture the maximum number of molecules. At pressures between the pure viscous and molecular flow conditions there is a transitional flow region (Knudsen flow) where flow characteristics have properties of both. In this region the back diffusion rates become larger and viscous flow becomes less effective at preventing backstreaming as the pressure drops.

As pressure drops, backstreaming (basically a molecular flow phenomenon) becomes more of a problem for pumping oils. The prevention of backstreaming basically becomes that of stopping the back migration of oil molecules to the sample chamber while the system is in transitional or molecular flow conditions. Surface creep of pump oil along the surfaces of the vacuum system must also be stopped. The basic ways to do this includes traps, changing types of pump oils, operational methods, foreline pressure modification, or the use of dry vacuum pumps.

Once a vacuum system is contaminated with oil, all surfaces which are oily will become secondary sources of oil within the system. Installing a corrective measure on the primary source of oil will not correct any secondary sources in the system. These secondary sources must be removed by cleaning.

VACUUM PUMP OILS USED BY HITACHI

Hitachi microscopes come from Japan with Lion S oil in the diffusion pump and Neovac MR-100 in the rotary pump. Because of the cost and difficulty involved with obtaining large quantities of these oils from Japan, HII has for some time now used Santovac 5 D.P. Oil and TKO-19 (Kurt J. Lesker company) rotary pump oil as replacement pump fluids. Both of the D.P. fluids have vapor pressures in the 10-10 torr range but are very different materials. Santovac 5 is a polyphenylether fluid a with very high viscosity at 25 C and Lion S is a Alkyl-naphtalene with much lower viscosity. In addition Lion S oil works best if the cooling water in the DP does not rise above 25 C and Santovac prefers a slightly warmer cooling water temperature because of its extreme viscosity at 25 C or below.

The rotary pump oils have vapor pressures of about 10-6 mbar at 25 C. If the DP fluid becomes contaminated with a significant amount of rotary pump oil, then the RP oil will distill out of DP oil during pumping and the ultimate vacuum of the DP will be that obtainable by the RP oil and not that of the low vapor pressure DP oil! Thus keeping the DP oil free of RP oil becomes important in obtaining a low pressures and a clean vacuum. The oil contamination found in several Hitachi microscopes has been analyzed and identified as rotary pump oil and not diffusion pump oil.

There are two possible ways for this oil to reach the sample chamber. One is by distillation out of contaminated diffusion pump oil, and the other is by direct backstreaming through the roughing line of the sample chamber or the exchange chamber. Backstreaming takes place by either in the vapor phase by direct evaporation or by surface creep of the liquid film. We will discuss methods to stop the oil from reaching the sample chamber.

TRAPS AND BAFFLES

One of the oldest ways of preventing backstreaming is to use a trap or baffle to prevent the oil from backstreaming. The oil condenses into the trap which prevents it from moving further up into the system. The many types of traps. On electron microscopes traps are found between the sample chamber and sometimes between the diffusion pump and on the foreline to the rotary pump.

Traps have the disadvantage that eventually oil molecules will find their way past especially if the trap is not cleaned or renewed regularly. Some traps will have a longer hold down time than others but all traps need renewal regularly. The trapping material in foreline traps eventually becomes saturated with oil or water, and then it will act as simply as a secondary source for vapor migration into the system. Traps slow backstreaming but without renewal eventually all the surfaces of the vacuum system will be coated with oil, and the lowest pressure obtainable in the system will be determined by the vapor pressure of the most volatile oil in the system. Trapping materials must be constantly maintained to be effective. Maintenance consists of either baking the material, cleaning, or replacement.

If a foreline trap is not being effective at stopping oil contamination in a system, the first thing to do is to check the condition of the trap and check the oil installed in the vacuum pumps. If the trap is saturated with oil the trapping material should be replaced and the rotary pump checked to make sure it is contains clean oil of the correct type and that oil is not being sucked back into the foreline. If the type of oil in either pump is unknown it should be replaced.

A new or renewed foreline trap will not clean up a contaminated vacuum system. Oil from previously contaminated surfaces will continue to creep and evaporate within the vacuum system after the trap is installed. This oil will eventually be captured by the trap, but this is a process governed exponential decay and has a long half life. A system cleaning is recommended to remove oil contamination.

Foreline traps that require baking for renewal are not suited for most electron microscopes because during the bake cycle the released gases and oils must be prevented from condensing elsewhere in the foreline. This requires either that the whole foreline be baked and there be a bakeable valve to seal the system from the trap, or that dry nitrogen be bled through the trap during baking to carry the released gases to the pump. Foreline traps for electron microscopes should have a replaceable or cleanable filter material.

There are number of types of foreline traps available. We will discuss the common traps and their advantages and disadvantages.

Molecular Sieve Traps: These traps work by adsorbing oil into a molecular sieve material which is very porous and has a very large surface area. To work effectively the molecular sieve must be baked out under vacuum to remove accumulated gases, water, and oil. This baking must be done frequently maintain the effectiveness of the trap. These traps prefer to trap water vapor over oil, and when they are full of water they are no longer effective for oil. Hitachi does not recommend molecular sieve traps because they must be baked to be refreshed and because water vapor makes them ineffective.

Copper Gauze Traps: Copper gauze reacts chemically with the oil vapor to trap it. In addition the gauze provides a high surface area and an optically dense path against molecular flow. When the copper gauze turns green and gunky it should be cleaned or replaced. The use of a transparent trap housing such as the Varian 346 is recommended because the condition of the copper gauze can be monitored from the outside by the user. This type of housing should be mounted so that any oil droplets that pass will puddle in the bottom of the trap. Depending on the system this trap will require maintenance every 2 to 6 months. Replacement copper gauze is easily found in the form of copper pot scrubber pads in the super market. These should be cleaned with a solvent to remove oil before use.

Stainless Steel Gauze Traps: These traps work only by providing a high surface area and a optically dense path against molecular flow. It is most effective against simple aerosols rather than the oil vapors. Unlike the copper gauze they do not react chemically with the oil vapor. Therefore their capacity is less than that of the copper gauze trap. Therefore these traps will need maintenance every 1 to 2 months to clean or replace the gauze. This type is not recommended.

Micromaze Trap (Kurt J. Lesker Company): This is a very effective foreline trap that uses a proprietary ceramic sheet material with many pores and a high surface area and acts like a molecular sieve in its ability to trap large or polar molecules such as water and hydrocarbon oils. To be effective it must be baked at 200 C for several hours before use and when it needs to be refreshed. Because it is baked in-situ this trap is not recommended for most systems.

In the mid 1990s experiments by David Joy of the University of Tennessee suggested that a continuously heated Micromaze trap would stop backstreaming. A trapping device using this arrangement (heated micromaze) as then sold by Hitachi as an accessory to their electron microscopes users to solve the backstreaming problem. The additional heat makes in hard for molecules to be adsorbed on the micromaze surfaces and slows down the migration of oil through he trap. In the long run, however, this arrangement did not prove satisfactory. David Joy no longer recommends this device.

Activated Alumina: This material is very effective in preventing the back migration of oil into the system and in trapping water and acid vapors before they reach the pump. It has a long life and needs to be renewed only every six months. Hitachi used activated alumina traps in the forelines of Turbo Molecular Pumps (TMP) in the early 0s. For renewal the activated alumina is poured out and replaced. Activated Alumina also actively absorbs water and acid vapors which may interfere with obtaining a good vacuum. Wet alumina will not pump down properly and must be dried and regenerated. Wet alumina may be regenerated by baking at 200 C outside of the vacuum system in an oven. Activated Alumina is a caustic material and there are certain handling precautions.

Liquid Nitrogen: Liquid Nitrogen traps are very effective for freezing out oil vapor. But they require constant refilling to stay effective. LN2 traps are best used in short term pumping situations where the pump is turned off or valved off when the pumping is complete. For foreline use the a LN2 trap requires either a regular filling schedule of an automatic filling system.

Traps and baffles for the diffusion pumps: Hitachi microscopes with diffusion pumps are equipped with a water cooled baffle mounted over the diffusion pumps. This baffle is meant to condense diffusion pump oil and return it to the D.P. with a small decrease in pumping speed. On some Hitachi Microscopes there is an additional LN2 cryotrap mounted over the diffusion pump. These traps need constant refilling to keep them effective.

FORELINE PRESSURE

Backstreaming in the foreline of the diffusion pump is the result of the pressure in the foreline being so low that molecular flow is possible for the oil molecules against the flow of the pumped gasses. Higher foreline pressures bring viscous flow conditions which act to keep the R.P. oil in the pump. If a rotary pump is only used for roughing and not as a forepump, the pump can be simply turned off to prevent backstreaming when a rough vacuum is achieved. Where the R.P. is used as a forepump this is not possible so other methods must be used to keep the fore pressure up.

One way to achieve higher foreline pressure and viscous flow conditions is by the use of a Nitrogen leak into the foreline which adjusts the pressure to one which causes viscous flow but which is low enough for the D.P. to work against. The nitrogen leak may be either a calibrated leak or an adjustable leak valve. Dry nitrogen is used rather than air to avoid the introduction of oxygen or water vapor into the system. This method is used in many types of process equipment to reduce foreline backstreaming.

A detailed discussion of this technique was published by Michael Postek of NIST. An Approach to the Reduction of Hydrocarbon Backstreaming in Scanning Electron Microscopes, Scanning Vol. 18, 269-274 (1996). Reprints of this Article are available from XEI Scientific.

CROSSOVER PRESSURE

Related to the question of foreline pressure is that of crossover pressure. Crossover pressure is the pressure at which the roughing valve to the sample chamber is closed, and the DP valve to sample chamber is opened. The crossover pressure can be adjusted by field service engineers for most microscopes.

If the cross over pressure is too low, then the RP will be connected for too long to the chamber and conditions favorable to direct backstreaming of RP oil into the chamber will result. If the crossover pressure is too high, then when the main valve is opened into the DP from the chamber, the resulting turbulent flow will upset the upper most jets of the DP and DP oil can come into the chamber. For microscopes with a single pump for both roughing and backing the DP there is the additional problem that at high crossover pressures the contents of the roughing line can be sucked into the DP foreline to contaminate the DP after crossover.

Therefore you may make a small adjustment upwards of the crossover pressure to reduce the direct backstreaming of RP oil into the chamber during roughing, but you must exercise care not to raise the crossover pressure too high or the pumping action of the DP will be disrupted releasing oil from the DP to the chamber when the main valve opens.

ROTARY PUMP OIL SELECTION

Backstreaming from the RP can also be reduced by using a pump oil with a lower vapor pressure. In addition if the Rotary Pump oil and the diffusion pump oils are the same, the contamination problem of the DP oil by the RP oil is eliminated. The diffusion pump oil must have adequate lubrication properties to be used in rotary pumps. Highly refined hydrocarbon oils meet this requirement. Silicone oils such as those made by Dow Corning have very poor lubricating properties and should not be used in RPs. Other oils such as ethers, esters, and fluorinated hydrocarbons depend on individual characteristics.

Santovac 5 oil has a very high viscosity and should not be used in an RP.

Lion S oil can be used in a RP successfully and has a vapor pressure in the 10-10 mbar range. Unfortunately it is available in only small quantities (in 100 ml bottles) and has limited availability since it is imported from Japan.

Diff Oil 20 (Kurt J. Lesker Co.) has a vapor pressure of 10-7 mbar and is suitable for both mechanical pumps and diffusion pumps according to the distributor. It is of reasonable price and readily available. While it will not give as low a ultimate vacuum as uncontaminated Santovac 5 it should lower the overall pressure and oil backstreaming in most Hitachi DP pumped systems by an order of magnitude because RP oil contamination is eliminated.

If the DP and RP oils are replaced in this manner, the whole vacuum system (Pumps, valves, foreline, and chambers) must be cleaned of the old RP oil before replacing the pump oils. If you do not do this, the old oil will continue to evaporate from the surfaces of the vacuum hoses, valves, pumps and chambers and the contamination will continue.

DRY VACUUM PUMPS

One of the cleanest ways to eliminate the backstreaming problem is to use a dry vacuum pump. There are several types available, but the type compatible with roughing sample chambers and backing oil diffusion pumps is a molecular drag pump. A drag pump is similar to a turbo pump but is designed to operate at higher pressures than a TMP (Turbo Molecular Pump). Because there is no oil in the pump, there is no backstreaming and contamination. The diffusion pump is them able to reach a pressure limited by the pressure of the diffusion pump oil vapor pressure alone.

The pump approved by Hitachi in 1990 for use on electron microscopes is a Alcatel DRYTEL 30. The Drytel 30 is molecular drag pump backed by a diaphragm pump. For roughing the system the Drytel pumps is slower than the standard rotary pump. Some cusomers may find it unacceptably slow. HII has available a DRYTEL pump conversion package available containing this pump with a conversion kit for connecting it to Hitachi electron microscopes. This pump has been placed through extensive reliability tests to insure that it does not fail though many S.C. and S.E.C. roughing cycles on a SEM.

 The Alcatel MDP 5011 molecular drag pump used in the Drytel 30 may be backed by a rotary pump instead of the diaphragm pump. This will significantly improve roughing speeds. In this case the drag pump acts like an active trap on the foreline to stop backstreaming. Unlike a passive foreline trap it does not need the trap material renewed on a frequent schedule.

TURBO MOLECULAR PUMPS (TMP)

Turbo molecular pumps are the most expensive solution to the oil backstreaming problem. The TMP replaces the diffusion pump in the system. The Hitachi TMP package consists of a Seiki-Seiko mag-lev TMP with activated Alumina Traps for the foreline and roughing line. The TMP will prevent oil from backstreaming through it. But the activated Alumina traps are necessary to prevent oil backstreaming during the roughing cycles. Thus it is necessary to keep the traps fresh and maintained on a regular cycle in order to keep the vacuum system dry.

For the ultimate in clean vacuum a TMP should be used in combination with a Molecular drag pump for roughing. In this way there is no oil in the system to cause contamination.

CRYOPUMPS

Currently there are no cryopumps approved for use with Hitachi electron microscopes because of vibration or reliability problems. A suitable cryopump may exist but has not been submitted yet to Hitachi for test and approval.

Since a cryopump is a pure capture type pump which does not require continuous backing, on a cryopump equipped microscope, clean vacuum operation could be achieved by operating the rotary pump only during roughing operations and avoiding low pressure molecular flow conditions. A cryopumped system would require considerable change in the evacuation control logic and circuits to function properly.

SUMMARY- STOPPING BACKSTREAMING

There is no one method for reducing oil backstreaming in electron microscopes. The choice depends acceptable cost, maintenance level, and degree of vacuum cleanliness desired. The following methods are available for reducing oil backstreaming:

CHANGE CROSSOVER PRESSURE: Cost very low. No maintenance.

Vacuum quality - small improvement.

ACTIVATED ALUMINA TRAP: Cost low. Needs renewal every 6 months.

Vacuum quality - medium

COPPER GAUZE TRAP: Cost low. Needs renewal every 2 months

With transparent trap it can be visually checked. Vacuum quality - medium

CALIBRATED FORELINE LEAK: Cost low - medium. Need source of dry nitrogen. Low maintenance. Vacuum quality - medium.

CHANGE TO DIFF OIL 20 IN BOTH PUMPS: Cost low. Low maintenance. No cross contamination of oils. Vacuum quality - medium-high.

MOLECULAR DRAG PUMP: Cost medium - low. Acts as a active foreline trap to stop backstreaming. With rotary pump produces fast roughing but system is not totally oil free. Less maintenance than a passive trap. Generates some noise. Vacuum quality - high-medium.

DRYTEL 30 PUMP: Cost medium. Replaces rotary pump. Needs more time for roughing. Needs PM to replace diaphragm every 6 months. Generates some noise. Vacuum quality - high.

TURBOMOLECULAR PUMP: Cost high. Replaces diffusion pump. Low maintenance. Vacuum quality - high.

SYSTEM CLEANUP IS REQUIRED to make any of the above methods effective. Residual oil contamination in the vacuum system will act as a secondary source of oil contamination on samples even after primary corrective action is taken.